Determination of paeoniflorin by HPLC (High Performance Liquid Chromatography)

Jul 07, 2021 View: 2 Leave a message

Application: The method was used to determine the content of paeoniflorin in Paeonia lactiflora Pall extract by HPLC.

ef8232fcd845046b365a77d219c6eeb

Method Principle: The sample was extracted by ultrasonic extraction with a proper amount of dilute ethanol in a conical flask, cooled, shaken and filtered. 

The filtrate was injected into a high performance liquid chromatograph for chromatographic separation. The absorption value of paeoniflorin was detected at 

230nm wavelength by ultraviolet absorption detector, and its content was calculated.

Reagents:

1. Acetonitrile (chromatographic purity)

2. Phosphate

3. Ethanol

Instruments and equipments:

1. The instruments

1.1 High performance liquid chromatograph

1.2 The chromatographic column:Octadecyl bonded silica gel as filler (4.6mm×250mm)

1.3 Ultraviolet absorption detector

2. Chromatographic conditions

2.1 Mobile phase: acetonitrile 0.1% : phosphoric acid solution = 15:85

2.2 Detection wavelength: 230nm

Sample preparation:

1. Weighing and preparing test samples

Weigh about 10g of test sample, finely ground and evenly, and finely take 0.1g.

2. Preparation of reference solution

Paeoniflorin 1mg was accurately weighed and added with ethanol to make a reference solution of 0.1mg/mL.

3. Preparation of test solution

The test product was placed in a stopper conical flask, and 25ml diluted ethanol was added accurately.Mix, weigh, ultrasonic treatment for 30min, cool down, 

then weigh the weight, dilute ethanol to make up for the lost weight, shake well, filter, take the filtrate as the test solution.

Note: "Precise weighing" means that the weighing should be accurate to one thousandth of the weight taken."Precise measuring" means that the accuracy of 

measuring volume shall meet the accuracy requirements of the pipet of such volume in the CP standard.


Operation steps:

1. Standard curve drawing

Accurately weigh paeoniflorin solution 4,7,10,13,16 μL and inject into HPLC respectively. The peak area integral value of Paeoniflorin 

was determined at 230nm wavelength by ultraviolet absorption detector. The standard curve was drawn by using the peak area integral value of injection volume 

regression.

2. Test sample determination

Paeoniflorin was measured at 230nm by ultraviolet absorption detector. The content of Paeoniflorin was calculated by external standard method.